ClearanceBrief
Recall Z-1954-2021 ยท DOF Lead, Atomic Absorption

Magellan Diagnostics, Inc.

LeadCare Ultra Blood Lead Test Kit Catalog Number 70-8098

DateJul 1, 2021 (initiated May 7, 2021)
StatusTerminated
ReasonMagellan has received reports that control tests of either the Low-Control and/or the High-Control generated a low result. The issue may cause the underestimation of blood lead levels in patient samples.
Root causeNonconforming Material/Component
ActionOn May 17, 2021, Magellan issued an Urgent Medical Device Removal notice to customers and distributors via letter notifying them Magellan has identified an ongoing issue with testing of the controls included in specified lots of LeadCareII Blood Lead Test Kits, LeadCare Plus Blood Lead Test Kits, and LeadCare Ultra Blood Lead Test Kits. REQUIRED ACTIONS: LeadCare blood lead test kits, Catalog Numbers: 70-6762, 82-0004, and 70-8098 - Per laboratory policies and procedures, laboratories should evaluate patient test results that were generated with the impacted lots. - Magellan is recommending
Quantity441 kits (192 tests/kit = 84,672 tests)
DistributionWorldwide distribution. US nationwide including Puerto Rico; Canada, Vietnam, India, Italy, Peru, New Zealand, Australia, Hong Kong, and Philippines.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K123563LEADCARE (R) ULTRA (TM) BLOOD LEAD TESTING SYSTEM
Magellan Diagnostics, Aug 20, 2013