Recall Z-1838-2021 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayStation (Treatment Planning System) Model Number RayStation is 6.0.0.24 , 6.1.1.2 , 6.2.0.7 , 6.3.0.6 , 7.0.0.19 , 8.0.0.6 , 8.0.1.10 , 8.1.0.47 , 8.1.1.8 , 8.1.2.5 , 9.0.0.113 , 9.1.0.933 , 9.2.0.483 , 10.0.0.1154 , 10.0.1.52 or 10.1.0.613 .

DateApr 21, 2021
StatusTerminated
ReasonEvaluation of a clinical goal for the voxelwise worst dose distributions may be misleading after editing the clinical goal. It also describes an issue with contouring when the spacing between slices is d 1 mm, present in RayStation/RayPlan versions from 6 up to 10A.
Root causeUnder Investigation by firm
ActionOn May 6, 2021, Raysearch issued an Urgent Medical Device Correction notice to customers via letter notifying them an issue was found with the voxelwise worst clinical goals in the Robust evaluation module for specified software versions of RayStation. Evaluation of a clinical goal for the voxelwise worst dose distributions may be misleading after editing the clinical goal. Actions to be taken by the user Voxelwise worst clinical goal in Robust evaluation - The user is advised to be careful when editing clinical goals in the Robust evaluation module and when evaluating the clinical goals for
QuantityDomestic: 13,722 systems
DistributionWorldwide distribution - US Nationwide distribution in the states of AK, AL, AZ, CA, CT, DC, DE, FL, GA, HI, ID, IL, LA, MA, MD, ME, MI, MO, MS, MT, NC, NJ, NM, NY, OH, OK, OR, PA, RI, SC, TN, TX, UT, VA, VT, WA, WI and the countries of Algeria, Australia, Austria, Bahrain, Belgium, Canada, Colombia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200569RayStation
RaySearch Laboratories AB (PUBL), Sep 24, 2020