Recall Z-1906-2021 ยท QKO Reagent, Coronavirus Serological

Beckman Coulter Inc.

Access SARS-CoV-2 lgG II Calibrator, REF C69058, IVD, CE UDI: (01)15099590742751

DateMay 7, 2021
StatusTerminated
ReasonSARS-CoV-2 lgG II numerical results may potentially be multiplied by a factor of 1000 on systems running with assay protocol file (APF) and access assay file (AAF) versions resulting in falsely elevated numerical values.
Root causeSoftware Design Change
ActionOn May 7, 2021, Beckman Coulter issued an "URGENT MEDICAL DEVICE RECALL" to affected consignees. The notification was emailed and sent via First Class Mail. In addition to informing consignees about the recall, the firm ask consignees to take the following actions: 1. Do not install any of the APF/AAF versions listed in the table. 2. Determine your current APF/AAF version. " From the Main Menu: select Configure F8 System Setup F1 System Revisions F1. " Verify the APF/AAF version displayed on the screen. 3 If the APF/AAF version is listed in the table: " Stop using the Access SARS-CoV-2 I
Quantity3,843 units
DistributionUS: Alabama, Alaska, Arizona, Arkansas, California, Florida, Georgia, Illinois, Iowa, Kentucky, Louisiana, Maine, Maryland, Massachusetts, Michigan, Mississippi, Missouri, New Hampshire, New Jersey, New York, North Dakota, Ohio, Oklahoma, Pennsylvania, Puerto Rico, Tennessee, Texas, Virginia, West V
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.