Recall Z-1775-2021 ยท LLZ System, Image Processing, Radiological

Fujifilm Medical Systems U.S.A., Inc.

Synapse Cardiovascular (Synapse CV). Software Versions: 6.0 to 6.2.1 The Synapse Cardiovascular image is a management and reporting system

DateApr 30, 2021
StatusTerminated
ReasonSoftware versions 6.0.4 to 6.2.1 using Advanced Reporting -possibility that a previously assigned internal patient database ID can be reused for a new patient. It was discovered that this value does not increase when a patient merge activity is executed in a specific sequence
Root causeSoftware design
ActionFujifilm issue consignee notifications via email on April 30, 2021. Letter states reason for recall, health risk and action to take: ACTIONS TO BE TAKEN BY CUSTOMER/END USER FUJIFILM will offer an upgrade to Synapse CV 6.2.2 to resolve this issue. FUJIFILM recommends the following actions: 1) Complete and return the Field Action Verification Form attached on page 3 of this communication. Providing the response with the information requested is essential for ensuring appropriate action is taken. 2) Contact Customer Service at 1-888-FUJIMED (1-888-385-4633) to open a ticket for the upgrade. Unt
Quantity44 US and 4 OUS
DistributionWorldwide distribution - US Nationwide distribution and the countries of Canada, Bermuda, Israel, Hadassah Medical Center.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K151859Synapse Cardiovascular
Fujifilm Medical Systems U.S.A, Inc., Aug 12, 2015