Recall Z-1955-2021 ยท GCX Apparatus, Suction, Operating-Room, Wall Vacuum Powered

Synaptive Medical Inc

Synaptive Trackable Suction Set Standard and Malleable

DateApr 23, 2021
StatusTerminated
ReasonDue to visible burrs/metal filings affixed along the suction tube inner perimeter.
Root causeVendor change control
ActionOn April 23, 2021 the firm sent a letter to its consignees with the following instructions: 1. Quarantine and discontinue use of all Synaptive Trackable Suction devices. 2. Contact Synaptive Recall Support at 1.866.853.8001 or by email at RecallSupport@synaptivemedical.com. Customer service will schedule an on-site inspection of each device. If this is not possible, arrangements will be made to have all impacted units returned to Synaptive. 3. Once on-site inspection or return of your stock has been arranged, please complete and return the attached Acknowledgement and Receipt Form. All forms s
Quantity1291 units
DistributionDomestic Distribution: IN, SC, MD, OR, TX ,LA, MI, NY, NV, FL, VA, CA,WI, PA and NC. International Distribution: Australia, Canada, Belgium, Netherlands, Singapore, Switzerland, Thailand, UK, Vietnam