Recall Z-1773-2021 ยท LLZ System, Image Processing, Radiological

CHANGE HEALTHCARE CANADA COMPANY

Change Healthcare Enterprise Viewer - Product Usage: intended to be used with off-the-shelf hardware for the 2D and 3D display of DICOM and non-DICOM medical images, reports, and multimedia content.

DateApr 19, 2021
StatusOpen, Classified
ReasonA software defect was identified where the Image Styles defined by presentation states from CHRS are not displayed in CHEV.
Root causeSoftware design
ActionOn 04/19/21 the firm sent a letter to its consignees with the following instructions: Immediate actions to take to reduce the risk to patient Prior to treating patients, always confirm the correct patient anatomy by reviewing the laterality markers applied during image acquisition. When the issue is detected, the clinician is able to manually adjust the Image Styles in CHEV to accommodate their viewing requirements and/or orient images to standard anatomical position. Product updates that will address the issue Change Healthcare is developing a software update to correct image display of ori
Quantity33 devices
DistributionUS Nationwide distribution in the states of AK, AR, AZ, CA, FL, HI, IA, IN, IL, KY, NE, NY, MA, MI, MS, OH, PA, TX,TN, WA, WI, WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181185Change Healthcare Enterprise Viewer