Recall Z-1766-2021 ยท NDN Cement, Bone, Vertebroplasty

Zavation

ZVplasty System, 20mm, Part# VCF-1020-2. For use in Orthopedic / spinal procedures.

DateMar 25, 2021
StatusTerminated
ReasonProducts distributed as sterile may not have been adequately sterilized
Root causeVendor change control
ActionZavation initially notified their consignees of the problem by email beginning on 03/24/2021 and requested they quarantine the product. The firm sent a subsequent email beginning on 04/13/2021 requesting the return of the product. The recall is being conducted to the user level.
Quantity24 units
DistributionUS nationwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.