Recall Z-1840-2021 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Nico Corp.

Myriad-LX Light Source-for delivery of light to the surgical field to enhance visualization of tissue during resection Part Number NN-7013 The Myriad-LX light source is a high-intensity 300W Xenon illuminator with a touch screen and digital dimming controls. The source is fitted with a proprietary connector to ensure usage is reserved for the Myriad-LX Illumination Fiber.

DateApr 30, 2021
StatusTerminated
Reason(1) Updated IFU for the LX Illumination Packs: Warning - Avoid contact between the distal opening of the delivery sleeve and tissue/debris within the surgical.field to prevent complete occlusion. Some surgical approaches (e.g., endonasal) may increase this potential. {f unintended reduction in light output occurs and/or occlusion of the sleeve is suspected, both the delivery sleeve and the fiber should be replaced (2) Removal for the LX Light Source potential for the distal tip of the sleeve/fiber assembly potential to burn healthy tissue resulting in minor to moderate injury to the patient
Root causeDevice Design
ActionThe Firm, Nico Corporation, issued "URGENT: MEDICAL DEVICE RECALL" Letter dated 4/30/21 via email on 4/30/21. Letter states reason for recall, health risk and action to take: ACTIONS TO BE TAKEN BY YOU Illumination Pack Field Correction 1. Determine the current number of LX Illumination Packs at your facility with the lot numbers referenced above. 2. Locate the electronic version of the updated Instructions-for-Use for LX Illumination Packs that was provided with this letter via email (document number LL-5697 03). 3. Print the updated Instructions-for-Use in a quantity sufficient to cover al
Quantity80 units
DistributionUS Nationwide Distribution
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K191599NICO Myriad NOVUS
Nico Corporation, Sep 13, 2019