Recall Z-1858-2021 ยท CEM Electrode, Ion Specific, Potassium

Radiometer Medical ApS

E3800 PC Units (spare parts)

DateApr 14, 2021
StatusOpen, Classified
ReasonDue to an issue with the internal clock of the analyzers with an e3800 PC unit installed.
Root causeDevice Design
ActionOn 04/14/2021, the firm initially sent an "Urgent: Medical Device Recall" Letter via email with a follow-up hardcopy via FedEx 2nd Day to customers inform them the Recalling Firm has become aware of a potential for the analyzers internal clocks may become incorrect. For analyzers which do not have an internal battery installed the incorrect time may be triggered in case power is abruptly removed from the analyzer, by e.g. toggling the power switch on the analyzer itself or at the wall outlet, or a power failure occurs on the mains supply. When the analyzer is switched on again, the analyzers
QuantityTotal 965 PC units World-wide (only 35 units in the U.S.)
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of AK, A L, AR, AZ, CA, CO, CT, DC, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, SC, SD, TN, TX, UT, VA, WA, WI and WV. The countries of Italy, Norway, Austr
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K092686ABL90 FLEX SERIES
Radiometer Medical Aps, Aug 6, 2010
K160153ABL90 FLEX PLUS
Radiometer Medical Aps, Nov 4, 2016