Recall Z-1864-2021 ยท OWB Interventional Fluoroscopic X-Ray System

GE Healthcare, LLC

GE Healthcare Innova IGS 3 (GTIN 00840682147378), Innova IGS 5 (GTIN 00840682124621), Innova IGS 6 (GTIN 00840682124614), Discovery IGS 7 (GTIN 00840682124638), and Discovery IGS 7 OR (GTIN00840682125888) interventional fluoroscopic x-ray systems.

DateJun 15, 2021 (initiated Apr 19, 2021)
StatusTerminated
ReasonThe IGS system can experience a single vertical line defect where the vertical line divides and horizontally shifts live monitor images into two unequal image parts within the monitor display screen.
Root causeComponent change control
ActionThe recalling firm issued letters dated 4/19/2021 via traceable mail on 4/20/21 and issued an additional letter dated 4/28/2021 to correct an omission of serial numbers (only sent to the affected consignees).
Quantity139 devices
DistributionWorldwide distribution - US Nationwide distribution in the states of CA, CO, FL, GA, KY, LA, MA, MO, MS, NC, NJ, NV, NY, PA, SC, TN, TX, VA, WA, and WI. Government distribution was made and potentially military distribution was made. The countries of Algeria, Argentina, Canada, China, Dominican Re
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181403Innova IGS 5, Innova IGS 6, Discovery IGS 7, Discovery IGS 7 OR