Recall Z-1675-2021 ยท JJE Analyzer, Chemistry (Photometric, Discrete), For Clinical Use

Roche Diagnostics Operations, Inc.

IRON GEN 2: ROCHE COBAS INTEGRA ANALYZER, REAGENT CASSETTES, Catalog#: 03183696122 - Product Usage: In vitro test for the quantitative determination of iron in human serum and plasma on Roche/Hitachi cobas c and cobas Integra systems.

DateApr 26, 2021
StatusOpen, Classified
ReasonRoche received several complaints regarding increased control recovery and discrepant elevated results for the Iron Gen.2 (IRON2) assay on the cobas c311, 501, and 502 analyzers and on the COBAS Integra 400 plus analyzer.
Root causeProcess control
ActionOn 04/26/2021, Roche issued an Urgent Medical Device Correction notice to customer via letter notifying them of issues regarding increased control recovery and discrepant elevated results for the Iron Gen.2 (IRON2) assay on the cobas c 311, 501, and 502 analyzers and on the COBAS INTEGRA 400 plus analyzer.
Quantity20,468 kits
DistributionUS Nationwide distribution in the states of AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, GU, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, UT, VA, VT, WA, WI, WV, WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K951595ROCHE COBAS INTEGRA ANALYZER, REAGENT CASSETTES
Roche Diagnostic Systems, Inc., Sep 8, 1995