Recall Z-1927-2021 ยท DQK Computer, Diagnostic, Programmable

GE Healthcare, LLC

Case Cardiac Assessment System for Exercise Testing and CardioSoft Diagnostic System Exercise Stress Testing ECG application

DateApr 19, 2021
StatusTerminated
ReasonIf a certain sequence of events occur, the pdf test report that belongs to one patient will appear in a different patient's record when viewed in an Electronic Medical Record (EMR) or a similar system.
Root causeSoftware design
ActionThe recalling firm issued letters dated 4/19/2021 beginning 4/19/2021. For systems that are on Windows 8.1 and above, the issue will be corrected through a software patch. For the units running on Microsoft Windows Server 2008-R2, Windows XP, and Windows 7, the letter instructs that the devices are not to be networked on these systems and that a software patch cannot be made available to these systems since the operating systems are not supported by Microsoft. A second cover letter and customer notification were issued dated 5/10/2021 updating the second safety instruction regarding the uni
Quantity11,295 devices (5,637 US; 5,658 OUS)
DistributionWorldwide distribution - US Nationwide. Government/military distribution was also made, and the countries of Algeria, Argentina, Australia, Austria, Azerbaijan, Bahamas, Bahrain, Bangladesh, Belgium, Bermuda, Brazil, Brunei Darussalam, Cambodia, Canada, Cayman Islands, Chile, China, Colombia, Costa
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.