Recall Z-1671-2021 ยท DSI Detector And Alarm, Arrhythmia

Philips North America Llc

IntelliVue MX40 Patient Monitor, Model Number: 865350 (US) 865352 (OUS) Software version number B.06.59 - Product Usage: Intended for monitoring and recording of, and to generate alarms for, multiple physiological parameters of adults and pediatrics in hospital environments and during transport inside hospitals.

DateFeb 2, 2021
StatusTerminated
ReasonManufacturing failed to follow process steps resulting in the distribution of potentially defective devices
Root causeEmployee error
ActionPhilips Healthcare conducted site visits on 2/02/21 by Field Service Engineer and provided the customer with RMA to return affected devices. Devices retuned affected device on RMA. Device Replacement provided.
Quantity3 units
DistributionWorldwide distribution - US Nationwide distribution in the states of IN, MN and the country of Italy.