Recall Z-1693-2021 ยท CBK Ventilator, Continuous, Facility Use

Draeger Medical, Inc.

Critical Care Ventilator, Catalog Number(s): 8422400: Babylog VN800 - Product Usage: used for treating patients who require temporary or longer term respiratory support for different medical reasons.

DateApr 16, 2021
StatusTerminated
ReasonThree separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800: 1. Restart of ventilation unit. 2. Incorrect FiO2 high and FiO2 low alarms. 3. SmartCare/PS (SC/PS) suspends weaning
Root causeSoftware design
ActionIn April, 2021, Draeger issued an Urgent Medical Device Correction notice to customer via letter notifying them that Draeger became aware of three separate and unrelated problems attributed to the software used in the Evita V600, Evita V800, Babylog VN600 and Babylog VN800:
Quantity12 units
DistributionUS Nationwide distribution in the states of PA, TX, MD, FL, TX, WA, MS, NY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.