Recall Z-1901-2021 ยท KDM Sigmoidoscope, Rigid, Non-Electrical

Adler MicroMed, Inc.

Sapi Med sterile disposable colorectal anoscopes and proctoscopes. REF/Product Description: A.4083/THE BEAK - Surgical proctoscope. Disposable & sterile. Transparent. Self light; A.4519/MPR Short - Multi Purpose Rectoscope. 190mm long. Sterile & Disposable; A.4086/EPOFLIER - Surgical kit made of a 34,3mm wide anoscope and an opened retractor named GULL. Disposable & sterile; A.4084/THE BEAK Diag

DateApr 23, 2021
StatusTerminated
ReasonDue to issue with product sterility. Sterilization service provider had falsified records.
Root causeProcess control
ActionOn 04/23/2021, the Recalling Firm communicated an URGENT: MEDICAL RECALL message via a telephone calls to customers and followed up email with a notification letter, informing them that the affected products that are labeled as sterile may not be sterile and the products could affect the sterile field and cause patient infections. Customers are instructed to: 1) Segregate and quarantine the affected products and wait for further instructions for disposition to follow, and 2) Complete and return the attached Acknowledgement and Receipt Form and indicate the quantities and location of the reca
Quantity1,858 devices
DistributionUS distribution: FL, IL, MA, and PA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K070913SELF-LIGHT DISPOSABLE ANOSCOPE
Sapimed S.P.A., Aug 10, 2007
K070915SAPIMED DISPOSABLE SIGMOIDOSCOPE
Sapimed S.P.A., Jul 6, 2007