ClearanceBrief
Recall Z-1939-2021 ยท FMK Single Use Only Blood Lancet With An Integral Sharps Injury Prevention Feature

Cardinal Health Inc.

ONETOUCH DELICA LANCETS 30G 100 COUNT Item Number: 5541610

DateMar 16, 2021
StatusOpen, Classified
ReasonDue to shipping delays the devices were exposed to temperature excursions of 31.9F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results.
Root causeEnvironmental control
ActionOn 03/16/2021, Cardinal Health issued an Urgent Medical Device Removal notice to customer via letter notifying them a recall is being initiated as a precautionary measure as a result of exposure to temperatures outside of the products labeling requirements. Actions Required 1. INSPECT your inventory for the affected product. 2. SEGREGATE and QUARANTINE all affected product. 3. RETURN the enclosed acknowledgment form via fax to 1-817-868-5362 or email to rxrecalls@inmar.com, even if you do not have any product on hand. 4. CONTACT Inmar to arrange for return and credit of any affected product:
Quantity2,315 boxes of 100 lancets
DistributionUS Nationwide distribution in the states of FL, GA, SC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.