Recall Z-1938-2021 ยท NBW System, Test, Blood Glucose, Over The Counter
Cardinal Health Inc.
ONETOUCH ULTRA TEST STRIPS 25 COUNT Item Number: 3510443
| Date | Mar 16, 2021 |
| Status | Open, Classified |
| Reason | Due to shipping delays the devices were exposed to temperature excursions of 31.9F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results. |
| Root cause | Environmental control |
| Action | On 03/16/2021, Cardinal Health issued an Urgent Medical Device Removal notice to customer via letter notifying them a recall is being initiated as a precautionary measure as a result of exposure to temperatures outside of the products labeling requirements. Actions Required 1. INSPECT your inventory for the affected product. 2. SEGREGATE and QUARANTINE all affected product. 3. RETURN the enclosed acknowledgment form via fax to 1-817-868-5362 or email to rxrecalls@inmar.com, even if you do not have any product on hand. 4. CONTACT Inmar to arrange for return and credit of any affected product: |
| Quantity | 144 boxes of 25 strips |
| Distribution | US Nationwide distribution in the states of FL, GA, SC. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K062195 | MODIFICATION TO ONE TOUCH ULTRA BLOOD GLUCOSE MONITORING SYSTEM Lifescan, Inc., Aug 16, 2006 |