Recall Z-1936-2021 ยท NBW System, Test, Blood Glucose, Over The Counter

Cardinal Health Inc.

ONETOUCH ULTRA BLUE TEST STRIPS 100 COUNT Item Number: 2990174

DateMar 16, 2021
StatusOpen, Classified
ReasonDue to shipping delays the devices were exposed to temperature excursions of 31.9F, intermittently over a 2 hour and 15 minute period of time which may cause inaccurate results.
Root causeEnvironmental control
ActionOn 03/16/2021, Cardinal Health issued an Urgent Medical Device Removal notice to customer via letter notifying them a recall is being initiated as a precautionary measure as a result of exposure to temperatures outside of the products labeling requirements. Actions Required 1. INSPECT your inventory for the affected product. 2. SEGREGATE and QUARANTINE all affected product. 3. RETURN the enclosed acknowledgment form via fax to 1-817-868-5362 or email to rxrecalls@inmar.com, even if you do not have any product on hand. 4. CONTACT Inmar to arrange for return and credit of any affected product:
Quantity5,356 boxes of 100 strips
DistributionUS Nationwide distribution in the states of FL, GA, SC.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K062195MODIFICATION TO ONE TOUCH ULTRA BLOOD GLUCOSE MONITORING SYSTEM
Lifescan, Inc., Aug 16, 2006