Recall Z-1630-2021 ยท MNT Continuous, Minimal Ventilatory Support, Facility Use Ventilator

Philips Respironics, Inc.

BiPAP A40, Brazil: Continuous ventilator, non-life-supporting. Material Number: BR1111169

DateApr 6, 2021
StatusTerminated
ReasonCertain devices were built with motor assemblies that could contain non-conforming plastic material. This could result in 1. Exposure to off-gassing of Volatile Organic Compounds not normally present creating a potential biosafety or toxicological hazard and 2. Sudden failure of the device causing a Ventilator Inoperative condition with the potential for asphyxia if not immediately identified and addressed by the care provider.
Root causeNonconforming Material/Component
ActionIn April, 2021, Philips issued an Urgent Medical Device Removal notice to customers via letter notifying them that Philips determined that certain devices were built with motor assemblies that could contain non-conforming plastic material. ACTION TO BE TAKEN BY CUSTOMER / USER 1. Locate all impacted devices listed above immediately STOP USE. All impacted devices must be removed from service. Replace with a non-impacted device if it is being used by a patient. 2. This notice needs to be passed on to all those who must be aware within your organization or to any organization where the impact
Quantity4 units
DistributionDistribution US nationwide to states of: LA, PA, WI, CO, WV, WA, CA, TX, MO, OK, FL, and TN. Brazil, China, and Japan