Recall Z-1543-2021 ยท FRN Pump, Infusion

STEP-HAR MEDICAL LLC

Alaris Pump Module, Model 8100, serviced/repaired by Step-Har Medical with an impacted bezel repair part

DateMay 10, 2021 (initiated Feb 26, 2021)
StatusTerminated
ReasonBezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Root causeComponent design/selection
ActionOn 02/26/2021, the one customer was notified of the recall via phone call and email. On 04/15/21, a Voluntary Field Safety Corrective Action notice were mailed. The firm asked their customer to respond so the bezel repair part could be replaced. Customers can contact the recalling firm at: 657-267-0054 or stdoan@stepharmedical.com
Quantity2
DistributionIA
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133532ALARIS SYSTEM WITH GUARDRAILS SUITE MX