The Biomed Guys
Alaris Infusion Pump Module, Model 8100, serviced/repaired by The Biomed Guys, with the impacted bezel repair part, TIPA-8100-4410 / Alaris 8100 Bezel.
| Date | Mar 25, 2021 |
| Status | Open, Classified |
| Reason | Bezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion. |
| Root cause | Under Investigation by firm |
| Action | The firm notified their consignees via email 3/25/2021 and multiple phone calls 3/25/2021, 3/26/2021 & 3/30/2021. Customers where also sent empty boxes with packing material to return their affected devices for inspection and remediation. |
| Quantity | 22 units |
| Distribution | Us distribution to the states of IN and TX. |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
| K133532 | ALARIS SYSTEM WITH GUARDRAILS SUITE MX Carefusion 303, Inc., Aug 21, 2014 |