Recall Z-1651-2021 ยท FRN Pump, Infusion

The Biomed Guys

Alaris Infusion Pump Module, Model 8100, serviced/repaired by The Biomed Guys, with the impacted bezel repair part, TIPA-8100-4410 / Alaris 8100 Bezel.

DateMar 25, 2021
StatusOpen, Classified
ReasonBezel repair part, not by the original manufacturer, used in the service and repair of Alaris Pump Module Model 8100, may experience cracking and/or separation of the bezel posts. Separation of one or more bezel posts could lead to free flow, over infusion, under infusion, or interruption of infusion.
Root causeUnder Investigation by firm
ActionThe firm notified their consignees via email 3/25/2021 and multiple phone calls 3/25/2021, 3/26/2021 & 3/30/2021. Customers where also sent empty boxes with packing material to return their affected devices for inspection and remediation.
Quantity22 units
DistributionUs distribution to the states of IN and TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133532ALARIS SYSTEM WITH GUARDRAILS SUITE MX
Carefusion 303, Inc., Aug 21, 2014