Recall Z-1563-2021 ยท JIX Calibrator, Multi-Analyte Mixture

Randox Laboratories Ltd.

Calibration Serum Level 3 (CAL 3) - multi-analyte mixture calibrator. Catalogue Number: CAL2351

DateMar 29, 2021
StatusTerminated
ReasonRandox can confirm a positive bias for Total Bilirubin and Direct Bilirubin of up to +15% across the assay ranges when using calibrator CAL2351, lot 1162UE. The positive bias has been attributed to bilirubin instability in the calibrator and is evident in both Quality Control (QC) and Patient samples.
Root causeNonconforming Material/Component
ActionOn 03/29/2021, Randox issued an Urgent Medical Device Correction notice to customer via letter notifying them, Randox can confirm a positive bias for Total Bilirubin and Direct Bilirubin of up to +15% across the assay ranges when using calibrator Calibration Serum Level 3. Action to be taken: -Discuss the contents of this notice with your Medical Director. -Complete and return the response form 12187-QA to technical.services@ randox.com within five working days. For questions or concerns contact Randox Technical Services.
Quantity98 kits
DistributionDomestic: IN, MA, CA, NJ, TX, NE, NY, VA, MD, MN, OR, IL, CT, DC, MS. Foreign: Albania, Argentina, Australia, Bangladesh, Bolivia, Bosnia & Herzegovina, Cameroon, Canada, Colombia, Cyprus, Denmark, Egypt, Estonia, Fiji, France, Germany, Greece, Hong Kong, Iraq, Italy, Ireland, Jordan, Kazakhstan, Ke
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K053153RANDOX CALIBRATION SERUM
Randox Laboratories, Ltd., Jan 13, 2006