Recall Z-1606-2021 ยท MDS Sensor, Glucose, Invasive

Medtronic Inc.

Medtronic REF: MMT-7020C Guardian Sensor (3) kit for the following packaging: GS3 SEMI 5PK OUS/ GTIN: 20763000192086, 5PK SF NEW GS3 OUS/GTIN: 20763000358451 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes. *Product not distributed within US.

DateFeb 26, 2021
StatusOpen, Classified
ReasonUse of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
Root causeUnder Investigation by firm
ActionOn February 26, 2021, Medtronic posted the "Urgent Recall Correction" notification to their website. On March 2, 2201, Medtronic provided U.S HCPs, Patients, Distributors and globally affected Clinical trial sites with a copy of the "Urgent Recall Correction" Notification. On March 8, 2021, Medtronic provided OUS patients, HCP and distributions with a copy of the "Urgent Recall Correction" notification Communication was provided to users via: email, fax, and/or mail. In addition to informing affected users about the correction, Medtronic ask consignees to take the following actions:
Quantity497,747 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov