ClearanceBrief
Recall Z-1604-2021 ยท MDS Sensor, Glucose, Invasive

Medtronic Inc.

Medtronic REF: MMT-7020LA, Guardian Sensor (3) for the following packaging: GS3 5PK US/ GTIN: 20763000336985 Product Usage: intended for use with Medtronic Diabetes (Medtronic) glucose sensing systems to continuously monitor glucose levels in persons with diabetes.

DateFeb 26, 2021
StatusOpen, Classified
ReasonUse of the continuous glucose monitoring system while actively taking hydroxyurea, an anti-neoplastic drug could result in inaccurate elevated sensor glucose readings.
Root causeUnder Investigation by firm
ActionOn February 26, 2021, Medtronic posted the "Urgent Recall Correction" notification to their website. On March 2, 2201, Medtronic provided U.S HCPs, Patients, Distributors and globally affected Clinical trial sites with a copy of the "Urgent Recall Correction" Notification. On March 8, 2021, Medtronic provided OUS patients, HCP and distributions with a copy of the "Urgent Recall Correction" notification Communication was provided to users via: email, fax, and/or mail. In addition to informing affected users about the correction, Medtronic ask consignees to take the following actions:
Quantity37,228 units
DistributionWorldwide distribution.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.