Recall Z-1490-2021 ยท KOH Instrument, Manual, General Obstetric-Gynecologic

Boston Scientific Corporation

Colpassist Vaginal Positioning Device - Product Usage: intended for use in general gynecological surgery to assist in the positioning and manipulation of the vagina during the procedure. Material No. M0068318210

DateApr 15, 2021
StatusOpen, Classified
ReasonPotential for pinholes (sterile barrier breach) on the edge of the pouch to potentially render the device non-sterile, and may result in a post-operative infection
Root causePackaging process control
ActionBoston Scientific issued Urgent Medical Device Product Removal on 4/15/21 via FedX. Letter states reason for recall, health risk and action to take; Individually packaged Upsylon Y Mesh (UPN M0068318200) is not impacted by this removal. This removal affects only the UPNs and batches listed below and in Attachment 1. No other UPNs or batches are impacted by this removal. Please cease further distribution or use of any remaining product affected by this removal immediately. If you identify any product from the affected lots/batches within your inventory, please segregate the product immediately
Quantity4,549 WW
DistributionWorldwide distribution - US Nationwide distribution and the countries of Algeria, Brazil, Colombia, Estonia, Germany, Hong Kong, Australia, Canada, Costa Rica, Finland, Great Britain, Italy, Austria, Chile, Czech Republic, France, Greece, Jordan, Belgium, Lebanon, Mauritius, Mexico, Morocco, Namibia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.