ClearanceBrief
Recall Z-1595-2021 ยท MMI Immunoassay Method, Troponin Subunit

Roche Diagnostics Operations, Inc.

Elecsys Troponin I STAT Immunoassay: Catalog# 05094798160 - Product Usage: intended to aid in the diagnosis of myocardial infarction.

DateApr 7, 2021
StatusOpen, Classified
ReasonRecall reason is possible sporadic false negative conventional troponin results. Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte is measured immediately before the Troponin I analyte on the same module, a decreased recovery on Troponin I was observed.
Root causeDevice Design
ActionOn 04/07/2021, Roche issued Urgent Medical Device Correction via fax notifying customers that Roche has confirmed an issue with running Elecsys Troponin I/Troponin I STAT and Elecsys proBNP II/ proBNP II STAT on the cobas e 601 and 602 modules. When the proBNP II analyte was measured immediately before the Troponin I analyte on the same module, a decreased recovery of Troponin I was observed.
Quantity10, 489 kits
DistributionUS Nationwide distribution in the states of AL, AZ, CA, CT, FL, GA, ID, IL, IN,KS, KY, MD, ME, MO, NC, NE, NJ, NM, NY, OH, OK, SC, SD, TN, TX, VA, WA, WY.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K082699ELECSYS TROPONIN I AND TROPONIN I STAT TEST SYSTEMS
Roche Diagnostics, Aug 19, 2009