Recall Z-1565-2021 ยท FPA Set, Administration, Intravascular

ICU Medical, Inc.

REF CH3397 Oncology Kit w/Spiros, Red Cap, Bag Spike w/Clave Additive Port LOT 5118610 - Product Usage: for use in the preparation and patient administration of cytotoxic medications.

DateMar 16, 2021
StatusOpen, Classified
ReasonDue to molding defect, there is a potential that certain lots of spinning and non-spinning Sprios male luers may exhibit small amounts of leakage in the inactivated position.
Root causeComponent design/selection
ActionOn 03/16/2021, the firm sent an "URGENT: MEDICAL DEVICE RECALL " Letter via Stericycle, Inc. informing customers that it has identified the potential for certain lots of Spiros to exhibit small amounts of leaks due to a molding defect. Customer are instructed to: 1) Discontinue the use and distribution of the affected product immediately. Check their inventory and quarantine all affected product at their facility. 2) Inform potential users of the product in their organization of this notification and complete the attached response form. Return the completed response form to the fax number or
DistributionWorldwide distribution - U.S. Nationwide distribution in the states of AK, AL, AZ, CA, CO, CT, DC, FL, GA, ID, IL, IN, KS, KY, LA, MA, MD, MI, MN, MO, MS, MT, NC, ND, NE, NJ, NV, NY, OH, OK, OR, PA, RI, SD, TN, TX, UT, VA and WA. The countries of Belgium, Canada, France, Italy, South Africa, Spain,
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081361CHEMOCLAVE CYTOTOXIC MEDICATION PREPARATION AND DELIVERY SYSTEM, MODEL, CH-XXXX
Icu Medical, Inc., Aug 28, 2008