ClearanceBrief
Recall Z-1593-2021 ยท FGE Stents, Drains And Dilators For The Biliary Ducts

W.L. Gore & Associates, Inc.

GORE VIBIL Biliary Endoprosthesis: stents, drains and dilators for the biliary ducts. Catalogue numbers: VH1008040, VH1010040, VN1010040, VN0808200 - Product Usage: intended for palliation of malignant strictures in the biliary tree.

DateApr 5, 2021
StatusTerminated
ReasonThe Gore post market surveillance team received a complaint that a GORE VIABIL Biliary Endoprosthesis appeared shorter than the labeled length.
Root causeProcess change control
ActionOn 04/08/2021, Gore issued an Urgent Medical Device Removal notice to customer via letter notifying them that a Gore Viabil Biliary Endoprosthesis 10 mm x 10 cm device appeared shorter than the labeled length.
Quantity61 devices
DistributionWorldwide distribution - US Nationwide distribution in the states of AZ, CA, CO, CT, FL, GA, HI, IL, KS, MD, MN, NE, NY, OK, RI, TX, VA and the countries of Portugal, Italy.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K081184GORE VIABIL BILILARY ENDOPROSTHESIS
W.L. Gore & Associates, Inc. Medical Products Div., Jul 22, 2008