ClearanceBrief
Recall Z-1484-2021 ยท DPS Electrocardiograph

Welch Allyn Inc Mortara

Surveyor S12/S19 Bedside Patient Monitor, with AM12M Acquisition Module accessory. Impacted software version of the AM12M is version 1.5.1 - Product Usage: indicated for use on adult and pediatric populations.

DateMar 31, 2021
StatusTerminated
ReasonDevices do not meet IEC 60601-2-27 requirements as labeled.
Root causeProcess control
ActionRecall notification letters were mailed to customers notifying them of the potential issue with recommended actions beginning April 7, 2021. The notification letter includes an acknowledgement form that must be returned to the recalling firm. Once the customer confirms the impacted units are in possession, the recalling firm will arrange for an exchange AM12M to be provided. The recalled unit will be returned to the recalling firm's service center for processing.
Quantity213 units
DistributionUS Nationwide distribution in the states of CA, CO, FL, GA, IN, LA, NC, NJ, NV, NY, OH, OK, TX, UT, and VA.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173765Surveyor Patient Monitor
Mortara Instrument, Inc., Sep 4, 2018