Recall Z-1436-2021 ยท GXP Methyl Methacrylate For Cranioplasty

Stryker Leibinger GmbH & Co. KG

DirectInject consists of a sterile dual paste system which is calcium phosphate based. Upon injection through the Mixer-Cannula, the two pastes form a cement paste which is moldable. The injected cement paste will harden under normal body conditions to form hydroxyapatite.

DateApr 1, 2021
StatusTerminated
ReasonInability for the user to inject the paste from the syringe into the target location.
Root causeUnder Investigation by firm
ActionOn 04/01/2021, Stryker issued an Urgent Medical Device Removal notice via letter informing customers that users have reported the inability to push the paste through the Mixer-Cannula which prohibits the injection into a bone void.
Quantity321 units
DistributionWorldwide distribution - US Nationwide distribution and the countries of Germany, Italy, France, Australia, Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K143661DirectInject
Stryker, Sep 2, 2015