Recall Z-1518-2021 ยท EME Punch, Biopsy, Surgical

Omnia SRL

Omnia Soft Tissue Punch is a dental instrument designed to remove with precision a circular section of oral soft tissue (e.g., gum, palatal mucosa) during dental surgery, typically to allow access to a site for drilling and dental/orthodontic implant placement without the use of a scalpel. Model: 32.Z2000.00, 32.Z2002.00, 32.Z2004.00, 32.Z2006.00, 32.Z2007.00, 32.Z2008.00, 32.Z2009.00, 32.Z2020.0

DateApr 29, 2021 (initiated Mar 30, 2021)
StatusTerminated
ReasonOmnia SRL received a notification from the contract manufacturer informing them of inadequate sterilization processes which have been taking place at their contracted sterilizing facility since 2016.
Root causeVendor change control
ActionOn 03/30/2021, Omnia issued an Urgent Medical Device Removal notice via letter to customers notifying them Omnia has been made aware of a situation at a contract sterilization facility regarding inadequate or inappropriate sterilization processes being followed by employees of the contract sterilizer.
Quantity276,901 units
DistributionDomestic: AK, IL, PA, OK, TX, UT, NY, CA. Foreign: Australia, Austria, Belgium, Bulgaria, Denmark, Dominican Republic, Estonia, France, Germany, Greece, Hungary, Ireland, Italy, Lebanon, Netherlands, New Zealand, Poland, Russia, Slovenia, Spain, Sweden, Switzerland, United Kingdom, Uzbekistan, Canad