ClearanceBrief
Recall Z-1504-2021 ยท ITI Wheelchair, Powered

Merits Holdings Co. dba Merits Health Products, Inc.

Power Wheelchair

DateApr 1, 2021
StatusOpen, Classified
ReasonSeat baseplate pin has broken in the field.
Root causeProcess change control
ActionA Recall notification letter titled, "MEDICAL DEVICE FIELD CORRECTION" was sent to consignees on 04/07/2021 via email and by certified mail on 04/16/2021. The firm stated they made phone call notifications on 04/01/2021. The letter instructs the consignee to Contact impacted customer using a personalized customer letter (template provided). The notification can be sent by mail, email or phone. (IF by phone, please keep a log of call. Date/Time/Result); replace the old seat plate with new one provided using the SOP provided. The field correction should take approximately 15 minutes to complete.
Quantity3054 units
DistributionDistributed nationwide to AL, AK, AZ, AR, CA, CO, CT, FL, GA, ID, IL, IN, IO, KS, KY, LA, MN, MD, MA, MI, MN, MS, MO, MT, NE, NV, NH, NJ, NM, NY, NC, ND, OH, OK, OR, PA, PR, SC, TN, TX, UT, VT, VA, WA, WV, WI and internationally to Canada
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K011513MP3U, POWER WHEELCHAIR
Merits Health Products., Inc., Aug 10, 2001
K011687MP-3C POWER BASE CHAIR, MODEL MP3C
Merits Health Products., Inc., Aug 3, 2001
K012802MP3CF ELECTRIC WHEELCHAIR
Merits Health Products., Inc., Sep 18, 2001
K122357MERITS MODEL P322FWD POWERED WHEELCHAIR
Merits Health Products Co., Ltd., Jan 9, 2013
K993032MODIFICATION TO MP-3 POWER BASE CHAIR
Merits Health Products., Inc., Nov 5, 1999