Recall Z-1495-2021 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Artis zee ceiling-interventional Fluoroscopic X-Ray System Model Number: 10094137

DateMar 26, 2021
StatusOpen, Classified
ReasonSome Artis zee ceiling systems show an increased abrasion of the cabling at the cable outlet of the inner C-arm, the limited functionality may not be sufficient to continue with treatment as planned. This may result in a situation where it is necessary to cancel clinical treatment or to continue treatment on an alternative system.
Root causeDevice Design
ActionSiemens issued Urgent Medical Device Correction letter dated March 26, 2021. Letter states reason for recall, health risk and action to take: Siemens will inspect the potentially affected area and will install an additional cover to improve the cable guidance and prevent further cable harness damage via Update Instruction AX008/20/S. If damage caused by abrasion is too severe, the cable harness will be replaced via Update Instruction AX011/20/S. Our service organization will contact you shortly to arrange a date to perform this corrective action. Please feel free to contact our service organiz
Quantity772 units U.S.
DistributionNationwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K181407Artis zee/zeego & Artis Q/Q.zen
Siemens Medical Solution USA, Inc., Aug 15, 2018