Recall Z-1433-2021 ยท JCW Prosthesis, Penis, Inflatable

Boston Scientific Corporation

The AMS 700 Inflatable Penile Prosthesis is a closed fluid-filled system containing saline. The system is made out of silicone and consists of 4 main components: a pump, two cylinders, a fluid reservoir and kink-resistant tubing that is used to connect the pump to the cylinders and fluid reservoir. By pumping the fluid within the system from the reservoir via the pump and into the cylinders, th

DateMar 26, 2021
StatusTerminated
ReasonAn AMS 700 100 mL Spherical Reservoir with InhibiZone was incorrectly labeled as a 100 mL Conceal Reservoir with InhibiZone.
Root causeProcess control
ActionOn 3/25/2021, BCS issued an Urgent Medical Device Removal notice via letter to customers notifying them certain serial numbers of AMS 700 Reservoirs may have been incorrectly labeled.
Quantity23 units
DistributionWorldwide distribution. US distribution to AR, AZ, CO, FL, GA, IA, MI, MO, NE, OR, PA, TN, TX, and UT. Belgium, France, Portugal