Recall Z-1427-2021 ยท MUJ System, Planning, Radiation Therapy Treatment

RAYSEARCH LABORATORIES AB

RayStation Model Number: (9A)9.0.0.113, 9B (9.1.0.933), 9B Service pack (9.2.0.483), 10A (10.0.0.1154), 10A Service pack (10.0.1.52), 10B (10.1.0.613)

DateMar 25, 2021
StatusOpen, Classified
ReasonFlags edited in RayCare after a patient has been checked in may not always be updated in RayTreat.
Root causeUnder Investigation by firm
ActionOn 03/25/2021, RaySearch Laboratories issued an Urgent Medical Device Correction notice via email to their customer notifying them of an issue found with the display of flags in RayTreat. Flags edited in RayCare after a patient has been checked in, may not be updated in RayTreat.
Quantity1 system
DistributionUS Nationwide in the state of TN.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K200569RayStation
RaySearch Laboratories AB (PUBL), Sep 24, 2020