Recall Z-1544-2021 ยท FRN Pump, Infusion

Infusion Pump Repair

Alaris Pump Module, Model 8100, serviced/repaired by Infusion Pump Repair with an impacted bezel repair part

DateMar 26, 2021
StatusOpen, Classified
ReasonBezel repair parts used, not by the original manufacturer, to service and repair infusion pump modules, may over time experience cracking or separation of the posts from the bezel. The separation of one or more bezel repair part posts may result in free flow, over infusion, under infusion or interruption of infusion.
Root causeComponent design/selection
ActionOn 03/26/2021, Voluntary Field Safety Corrective Action notices started being mailed to customers. The firm conducted further communications via email and phone. The firm asked customers with pumps serviced with the affected bezels to return the pump module units to the recalling firm so they could swap out the affected bezels or provide a refund. The firm asked customers to immediately share this field safety corrective action notification with all necessary parties within your organization to ensure that they are also aware of this action. Also, please complete and return the Customer Respo
Quantity15
DistributionCalifornia
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K133532ALARIS SYSTEM WITH GUARDRAILS SUITE MX
Carefusion 303, Inc., Aug 21, 2014