ClearanceBrief
Recall Z-1479-2021 ยท DRS Transducer, Blood-Pressure, Extravascular

Smiths Medical ASD Inc.

HIGH PRESSURE ROTATOR (Component) a. WITH MALE LUER LOCK, Model Number MX496HP b. WITH FEMALE LUER LOCK, Model Number MX497HP. component of extravascular blood-pressure Transducer.

DateApr 21, 2021 (initiated Mar 31, 2021)
StatusTerminated
ReasonSpecific lots of Medex High Pressure Adaptors, Rotators, Manifolds, Stopcocks and Tubing and Medex LogiCal Kits may have been manufactured with insufficient or incomplete welds. These are components for various cardiac kits including pressure monitoring devices and interventional imaging devices.
Root causeProcess control
ActionThe firm initiated the action by email on 03/31/2021. The Recall Notice instructs consignees to identify, and quarantine affected product in their possession. A list of affected product has been included on the Recall Notice for consignees to identify affected product. The consignees have been instructed to complete the response form, and return it to fieldactions@smiths-medical.com within 10 days of receipt. After the completed Recall Response Form has been submitted to fieldactions@smiths-medical.com, the consignee will be contacted to arrange return of any affected product in their possess
Quantity8150 units
DistributionDistribution US nationwide including Puerto Rico, Austria, Canada, Chile, Ecuador, Hong Kong, Mexico, Slovenia, Brazil, South Africa, Colombia, and Japan.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.