Recall Z-1503-2021 ยท PCL Enzyme Linked Immunoabsorbent Assay, Rubeola Igm

Clark Laboratories, Inc.

Trinity Biotech Captia Measles IgM Kit, Catalog No. 2326060 - Product Usage: intended for the qualitative detection of Measles igM antibodies in human serum of patients suspected of measles (rubeola) infection.

DateMar 17, 2021
StatusTerminated
ReasonPositive Control may run high out of its assigned ISR range and result in an invalid run and necessitate retesting
Root causeUnder Investigation by firm
ActionTrinity Biotech issued Product Notification Letter on March 17, 2021 stating reason for recall, health risk and action to take: Please note, if a laboratorys test resulted in a valid run, such as all kit Controls and Calibrator met the Quality Control Criteria as stated in the package insert, the patient results can be reported and a re-evaluation of patient results is not necessary. Customers are instructed to discontinue the use of any Measles IgM Lot #2326060-060 kits that they may have in their inventory. Please complete the attached fax back form informing us of how many of these kits you
Quantity80 kits
DistributionWorldwide distribution - US Nationwide distribution and the country of Ireland.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K140455CAPTIA MEASLES IGM
Trinity Biotech USA, May 22, 2014