Recall Z-1497-2021 ยท OWB Interventional Fluoroscopic X-Ray System

Siemens Medical Solutions USA, Inc

Interventional Fluoroscopic X-Ray System ARTIS icono biplane- 11327600 ARTIS icono floor- 11327700 ARTIS pheno- 10849000 ARTIS is a family of dedicated angiography systems developed for single and biplane diagnostic imaging and interventional procedures including, but not limited to, pediatric and obese patients.

DateMar 19, 2021
StatusTerminated
ReasonSiemens has become aware of a potential issue with software version VE20C. Planned procedures may have to be terminated and performed on an alternative x-ray system.
Root causeSoftware design
ActionOn 3/19/2021, Siemens Healthineers, issued an "URGENT: MEDICAL DEVICE CORRECTION" notice via letter dated 3/192021 to customers informing them of a possible hazard to patients, operators, or other persons and equipment due to a software error which affects ARTIS icono /pheno systems with software VE20C. The customer are instructed to do the following: The user should be aware of the limited system functionality in backup mode and bypass fluoroscopy operating mode (see also Operator Manual, Chapter 6.7: System Operating Modes). In some rare cases, the automatic recovery phase may not be succ
Quantity133 systems
DistributionWorldwide distribution: US (nationwide) including states of: CA, CO, DE, FL, IL, IN, KY, MD, MI, NC, NE, NJ, NY, OH, TN, TX, UT, WI, and OUS (countries) of: AT, BE, ES, IT, FI, CH, FR, GB, SE, DE, DK, IE and NL.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K193326ARTIS Icono
Siemens Medical Solutions USA, Inc., Apr 16, 2020
K201156ARTIS pheno (VE2)
Siemens Medical Solution USA, Inc., Jun 29, 2020