ClearanceBrief
Recall Z-1419-2021 ยท GDW Staple, Implantable

Covidien Llc

Covidien Tri-Staple 2.0 Reinforced Intelligent Reload 60 mm Medium/Thick-For use with Signia stapling system. Staple line reinforcement has applications in abdominal, gynecologic, pediatric and thoracic surgery for resection, transection of tissue and creation of anastomosis. It may be used for transection and resection of liver substance, hepatic vasculature and biliary structures, and for transe

DateMar 23, 2021
StatusTerminated
ReasonPotential for the pouch to contain an extra laminate layer, creating the potential for the device sterility to be compromised. The inclusion of an additional laminate layer in the package may also impact complete de-gassing of the sterilization gas, Ethylene Oxide and may result in a potentially increased risk for patient infection and/or potential exposure to Ethylene Oxide retained in the pouch
Root causeNonconforming Material/Component
ActionMedtronic notified consignees by letter delivered via Federal Express and/or certified mail beginning March 23, 2021. The letter states reason for recall, health risk and action to take: " Immediately quarantine and discontinue use of the affected item code and lot. " Return affected product. " Forward the information to anyone who they have distributed affected Covidien Tri-Staple" 2.0 Reinforced Intelligent Reloads. " Complete the Recalled Product Return Form Questions regarding this communication, please contact your Medtronic representative or Customer Service at 800-962-9888, option 2.
Quantity222 units
DistributionAL, AR, FL, IN, LA, MA , MN, MO, NJ, NY, OR, PA, TX
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173270Tri-Staple 2.0 Reinforced Reload
Covidien, Feb 20, 2018