Recall Z-1348-2021 ยท LLZ System, Image Processing, Radiological

Fujifilm Medical Systems U.S.A., Inc.

Synapse PACS - Radiological Image Processing System - Product Usage: intended for use, as a web based application, on an off-the-shelf PC meeting or exceeding minimum specifications and networked with FUJIFILM Synapse PACS Software (Server).

DateMar 2, 2021
StatusOpen, Classified
ReasonThe wrong patient information may be displayed in the viewer or PowerJacket.
Root causeSoftware design
ActionOn March 2, 2021, FUJIFILM Medical Systems U.S.A., Inc. (FUJIFILM) issued an Urgent Medical Device Recall notice for the voluntary recall of Synapse PACS versions 5.1 and higher via certified mail.
Quantity839 systems
DistributionWorldwide distribution - US Nationwide distribution including in the states of AZ, CA, CO, CT, DC, FL, GA, HI, IA, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, NE, NH, NJ, NV, NY, OH, OK, OR, PA, PR, RI, SC, SD, TN, TX, VA, VT, WA, WI, WV, WY and the countries of United Arab Emirates, Ang
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K160108Synapse PACS
Fujifilm Medical Systems U.S.A, Inc., Feb 3, 2016