Recall Z-1487-2021 ยท FRN Pump, Infusion

Baxter Healthcare Corporation

Spectrum IQ - Infusion Pump - Product Usage: intended to be used for the controlled administration of fluids.

DateMar 19, 2021
StatusTerminated
ReasonPotentially defective component in the AC power adapter which may fail over time.
Root causeDevice Design
ActionOn 3/19/2020 the firm sent a letter to its customers with the following instructions: 1. Operators may continue to use the Spectrum IQ Infusion Pumps until the correction is completed. If using an impacted pump to infuse critical medication such as vaso-active medications, ensure there is a back-up pump available in case the power adapter fails. 2. A local Baxter service representative will contact your facility to provide a list of affected serial numbers at your facility and to coordinate power adapter replacement. Your facility will be receiving this power adapter from Baxter at no charge.
Quantity408 units
DistributionWorldwide distribution - US Nationwide distribution in the states of AL, AR, CA, CT, FL, IA, IN, LA, MA, MD, ME, MN, MO, MS, MT, NC, NJ, NY, OH, PA, RI, SC, TN, TX and the country of Puerto Rico.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K173084Spectrum IQ Infusion System, Dose IQ Safety Software
Baxter Healthcare Corporation, May 11, 2018