Recall Z-1460-2021 ยท MRM Defibrillator, Implantable, Dual-Chamber

BIOTRONIK Inc

BIOTRONIK INVENTRA 7, VR-T DX DF-1, REF 399437, UDI: 04035479136321 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.

DateMar 8, 2021
StatusTerminated
ReasonThere is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Root causeDevice Design
ActionOn 3/8/21, BIOTRONIK issued a "Urgent Field Safety" recall notification to all affected consignees via UPS. In addition to informing consignees about the recall, the firm asked consignees to take the following actions: 1. To determine if a specific device is affected by this corrective action you can lookup the device at the following web address: www.biotronik.com/devicelookup. 2. Devices in stock: Do not implant any potentially affected devices. Local BIOTRONIK representatives will replace affected devices in hospital inventory. 3. Continue with the standard patient follow-up schedule. -