Recall Z-1451-2021 ยท MRM Defibrillator, Implantable, Dual-Chamber
BIOTRONIK Inc
BIOTRONIK ITREVIA 7, VR-T DF-1, REF 393040, UDI: 04035479129552 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.
| Date | Mar 8, 2021 |
| Status | Terminated |
| Reason | There is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds) |
| Root cause | Device Design |
| Action | On 3/8/21, BIOTRONIK issued a "Urgent Field Safety" recall notification to all affected consignees via UPS. In addition to informing consignees about the recall, the firm asked consignees to take the following actions: 1. To determine if a specific device is affected by this corrective action you can lookup the device at the following web address: www.biotronik.com/devicelookup. 2. Devices in stock: Do not implant any potentially affected devices. Local BIOTRONIK representatives will replace affected devices in hospital inventory. 3. Continue with the standard patient follow-up schedule. - |
| Quantity | 499 units |
| Distribution | US: Nationwide OUS: Worldwide |
| FDA record | Open on fda.gov |
510(k) clearances named in this recall
This recall record lists no 510(k) number.