ClearanceBrief
Recall Z-1446-2021 ยท MRM Defibrillator, Implantable, Dual-Chamber

BIOTRONIK Inc

BIOTRONIK Iperia 7, DR-T, DF-1, ProMRI REF 392409, UDI: 04035479128777 - Product Usage: intended to provide ventricular tachycardia pacing and ventricular defibrillation for automatic treatment of life-threatening ventricular arrhythmias.

DateMar 8, 2021
StatusTerminated
ReasonThere is a potential for premature battery depletion in Implantable Cardioverter Defibrillators and Cardiac Resynchronization Therapy Defibrillators (CRT'Ds)
Root causeDevice Design
ActionOn 3/8/21, BIOTRONIK issued a "Urgent Field Safety" recall notification to all affected consignees via UPS. In addition to informing consignees about the recall, the firm asked consignees to take the following actions: 1. To determine if a specific device is affected by this corrective action you can lookup the device at the following web address: www.biotronik.com/devicelookup. 2. Devices in stock: Do not implant any potentially affected devices. Local BIOTRONIK representatives will replace affected devices in hospital inventory. 3. Continue with the standard patient follow-up schedule. -
Quantity1652 UNITS
DistributionUS: Nationwide OUS: Worldwide
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.