Recall Z-1346-2021 ยท KPR System, X-Ray, Stationary

Siemens Medical Solutions USA, Inc

Ysio X.pree, Model Number: 11107464 - Product Usage: A Digital X-Ray System generate Xray images from the whole body including the skull, chest, abdomen, and extremities.

DateMar 10, 2021
StatusTerminated
ReasonA patient image was assigned to a different patient during image recovery process. In the unlikely event of interrupted connection to a MAX detector during initiated image recovery, instead of the currently acquired image, previously acquired image (potentially image of the previous patient) may be displayed to the user, potentially impact diagnosis of the patient examined when the connection was interrupted
Root causeDevice Design
ActionSiemens Medical Solutions USA, Inc. issued Customer Safety Advisory Notice to customers via XP051/20/S on 3/10/21. Letter states reason for recall, health risk and action to take: It is advised for the user to check the image displayed following image recovery to ensure that it is a newly acquired image of the current patient. If the operator detects the error, the image should be deleted, and Siemens Customer Service should be notified. Siemens Healthineers will provide a field software update VA10D for YSIO X.pree units that will eliminate the issue described. The software update will be d
Quantity12 units
DistributionUS Nationwide distribution in the states of FL, MN, OH, PA, TX.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K201670YSIO X.pree