ClearanceBrief
Recall Z-1530-2021 ยท NVZ Pulse Generator, Permanent, Implantable

St. Jude Medical, Cardian Rhythm Management Division

ASSURITY Pulse Generator REF PM**** SN ********* ST. JUDE MEDICAL Assurity and Endurity pacemakers are intended to provide single-chamber or dual-chamber bradycardia pacing.

DateMay 7, 2021 (initiated Mar 15, 2021)
StatusOpen, Classified
ReasonDue to a manufacturing issue associated with the mixing process of the epoxy used in the manufacture of pacemaker from years 2015 through 2018, the affected devices may malfunction caused by moisture ingress through the device header and feedthrough
Root causeProcess control
ActionOn 10/5/2021, the firm notified additional customer via a "Safety Notification Update" letter due to finding additional affected products after March 2021. The instructions from the 3/15/2021 original letter still applies. Issue of the update remains, the issue is caused by intermittent incomplete mixing of epoxy during manufacture, which may allow moisture ingress into the pulse generator header. As a result, the potential for affected clinical impact has included loss of telemetry/communication, reduced battery longevity, loss of pacing, and/or shortened duration between Elective Replaceme
Quantity242,945 pacemakers [updated 3/20/2023]
DistributionWorldwide distribution: U.S (nationwide) to states including.: AK, AL, AR, AZ, CA, CO, CT, DC, DE, FL, GA, HI, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MN, MO, MS, MT, NC, ND, NE, NH, NJ, NM, NV, NY, OH, OK, OR, PA, Puerto Rico, SC, SD, TN, TX, UT, VA, WA, WI, WV, and WY. O.U.S.( countries): [8
FDA recordOpen on fda.gov

510(k) clearances named in this recall

This recall record lists no 510(k) number.