Recall Z-1411-2021 ยท GEI Electrosurgical, Cutting & Coagulation & Accessories

Covidien

Covidien Surgiwand" II Suction and Irrigation Device: Electrosurgical, cutting & coagulation & accessories22 Product Number/CFN: 178083, 178093, 178094

DateMar 5, 2021
StatusOpen, Classified
ReasonInvestigations related to customer complaints received for particulate matter inside of the Surgiwand" II device tubing components identified the root cause as a damaged Y-Connector component which connects the tubing to the device. This damage was determined to generate the particulate matter observed and is created during the assembly process for connecting the tubing component to the Y-Connector.
Root causeProcess design
ActionOn 3/10/2021, Medtronic issued an Urgent Medical Device Recall notices to customers. All consignees were notified by letter delivered via Federal Express and/or certified mail. The letter informs customers of the recall of specific item codes and production lots of Covidien Surgiwand" II Suction and Irrigation Device following reports from customers of foreign particles in the device tubing.
Quantity59,828 devices
DistributionDomestic: AK, AL, AZ, CA, CO, CT, FL, GA, IA, ID, IL, IN, KS, KY, LA, MA, MD, ME, MI, MO, MS, NC, NE, NH, NJ, NM, NY, OH, OR, PA, SC, TX, UT, WA, WI, WV. Foreign: Canada, Japan, Taiwan, Australia, New Zealand, Indonesia, Thailand, Singapore, Malaysia, Philippines, Bahamas, Brazil, Chile, El Salvador
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K961771AUTO SUTURE SURGIWAND
United States Surgical, A Division of Tyco Healthc, Sep 25, 1996