Recall Z-1428-2021 ยท GCJ Laparoscope, General & Plastic Surgery

Stryker Corporation

Stryker, REF PC9001, PINPOINT Video Processor/Illuminator, Rx Only CE 0197, MD Medical Device, UDI: (01) 00858701006049 - Product Usage: used with both the SPY-PHI and PINPOINT systems, the intended use for both imaging systems are being documented here.

DateFeb 26, 2021
StatusTerminated
ReasonTheir is a potential that Video Processor/Illuminators may exhibit flickering, loss of image or unintended reboots.
Root causeComponent design/selection
ActionOn February 26, 2021 Stryker sent a "Urgent Medical Device Recall notification via mail to all affected consignees. In addition, to informing consignees about the reason for a recall, the firm ask consignees to take the following actions: 1. Immediately check inventory to locate the product listed on Attachment A and remove them from their point of use. 2. Return the enclosed business reply form by email to endorecall@stryker.com to confirm receipt of this notification/document product segregation. RESPONSE IS REQUIRED. a. If affected product is found, segregate the product and call Stryke
Quantity125 units
DistributionWorldwide distribution - US Nationwide distribution including in the states of CA, CO, FL, IL, MA MI, MS NC, NV, OK, PA, SC, TN, TX, VA, WA, WI and the countries of AE, CA, CN, CZ, DE, ES, HK, JP, KR, KW, MX, MY, PH, RO, SG, TR, TW.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K150956PINPOINT Endoscopic Fluorescence Imaging System
Novadaq Technologies, Inc., May 23, 2016