Recall Z-1373-2021 ยท LRG Instrument For Auto Reader & Interpretation Of Overnight Suscept. Systems

Beckman Coulter Inc.

MicroScan WalkAway-96 plus Reconditioned Instrument

DateSep 4, 2019
StatusTerminated
ReasonInstruments were manufactured with a shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure.
Root causeProcess control
ActionOn 09/04/2019, the firm sent an "URGENT MEDICAL DEVICE RECALL" Letter to its customers via electronic and/or first class postal mail informing them that an investigation revealed that some instruments were manufactured with shortened 30mL reagent draw straw which may result in reagents not dispensing and may lead to organism misidentification and/or Quality Control (QC) failure. Customers were also informed that when a reagent is not dispensed, there is the potential for any of the following to occur: -high probability misidentifications and reporting of erroneous results -panel reads are fla
Quantity9 systems
DistributionU.S.: CO, FL, IL, IN, KY, ME, MI, MS, PA, Puerto Rico, TN, TX, UT, VA and WA O.U.S.: India, Indonesia, Italy, Japan, Philippines, Poland, Russia, South Korea, Taiwan and Yemen.
FDA recordOpen on fda.gov

510(k) clearances named in this recall

K911400WALKAWAY TM 40
Baxter Healthcare Corp, Nov 19, 1991